About this work

What it captures
Who makes the device and who is assessing it, what the device is for, how it touches the body, whether it is powered, whether it contains software, any special characteristics (medicinal substances, tissues, nanomaterials, and so on), the Class I suffix modifiers (sterile, measuring, reusable), and the evidence and systems a manufacturer needs in place.
What it computes
The classification rules of MDR Annex VIII are applied to your answers and the highest class wins. You get the computed class, the final class (you may override it with a recorded reason), the conformity-assessment route, the rule that decided the class, a readiness verdict, and a list of flags with suggested actions. In vitro diagnostic devices are out of scope: they fall under the IVDR and the form tells you so.
Specification
Spec-driven: the living domain spec defines the data model, the classification engine and the flag rules. The same pure engine grades the wizard, the report and the dashboard.
Grounding
EU Regulation 2017/745 (MDR) Annex VIII, the UK Medical Devices Regulations 2002, MDCG 2019-11 guidance, and the ISO 14971, ISO 13485 and IEC 62304 standards. This is a simplified aid to a classification discussion, not a substitute for the regulation or for advice from a notified body.